    ABX Pentra XL 80 Hematology Analyzer

    High performance hematology analysis with integrated validation

    • CBC with 5-part Diff including flagging for immature cells
    • Advanced technologies for mixing and sample analysis ensuring accurate results
    • Processes 80 samples/hr and provides results in 60 seconds
    • Autoloader holds up to 100 samples with random continuous access
    • STAT mode for critical samples

    Features

    • 27 parameters
    • Whole blood micro-sampling (CBC: 35 µL, Plus Diff 53 µL) 
    • Cytochemistry, Impedance (real cell volume measurement) & Optical (analysis of the internal cellular structure by measuring light absorbance) - DHSS* Technology
    • Perfect homogenization of blood samples with reagents - MDSS** Technology
    • Automatic Sample Re-run
    • Reagents: Only 4 reagents plus cleaner
    • Perfect differentiation of the 5 WBC sub-populations with DHSS* Technology
    • 3 histograms for RBC, BAS/WBC and PLT together with the 5 DIFF Matrix
    • Basophils counted through specific channel
    • High resolution matrix includes the determination of 2 additional subpopulations (% and #): Atypical Lymphocytes (ALY***) and Large Immature Cells (LIC***)
    • Customized Dilution Ratio (CDR)
    • Integrated Workstation and Validation Station  


    * DHSS : Double Hydrodynamic Sequential System 
    ** MDSS : Multi-Distribution Sampling System 
    *** RUO parameters (Research Use Only)

    Specification


    Technical

    Weight 122 lb / 55 kg
    Power supply 100 V to 240 V (± 10%) - 50 Hz to 60 Hz
    Dimensions 32.3 x 22.4 x 21.5 in / 82 cm x 57 cm x 54 cm
    Printer Laser
    Throughput Up to 80 samples/hour
    Data Processing Colour LCD touch screen: 12 in
    PC onboard: Celeron 566 MHz
    Windows XP Prof.TM
    RAM (256 Mo), Hard disk (10 Go mini)
    Floppy disk & CD ROM reader
    RS 232C, 2 X USB1
    User defined flagging limits
    Transmit patient files & QC to LIS
    Uni-directional & bi-directional connections
    ASTM protocol inside
    Memory capacity 10 000 results + Graphics
    Quality Control Management 24 selectable QC files
    XB: 60 operator selectable files with statistics (20 results per file)
    With-in run
    Levey-Jennings graphs
    Logs Reagents, Quality Control, Calibration, Maintenance, Blank Cycle, Patients Data, Settings, Information, Technology Communication, Errors, by date.
    Patient reports Management Delta check
    Anteriority (Matrix, curves, data)
    Manual input

    Samples

    80 samples/hour (open or closed tube) 

    Parameters (27) WBC, NEU# & NEU%, LYM# & LYM%, MON# & MON%, EOS# & EOS%, BAS# & BAS%, ALY# & ALY%*, LIC# & LIC%*, RBC, HGB, HCT, MCV, MCH, MCHC, RDW CV, RDW SD, PLT, MPV, PCT*, PDW*.
    Graphs High-definition LMNE matrix
    PLT, RBC, WBC/BAS curves
    Blood Volume CBC mode: 35 µL
    CBC + DIFF mode: 53 µL
    Stability 48 hours post-draw stability
    Operating sample mode Autoloader 10 racks (100 tubes) capacity
    Sample mixing with end-over-end tube rotation
    Closed or open STAT tube
    Identification Alphanumeric. Bar-code reader (optional) and internal bar-code scanner

     * RUO parameters (Research Use Only)

    Results

    Parameters Precision (%CV) Linearity Limits Linearity Limits with CDR*
    WBC < 2.0 0-120 x 10³/µL 120-360 x 10³/µL
    RBC < 2.0 0-8 106/µL 0-8 X 106/µL
    HGB < 1.0 0-24 g/dl 0-24 g/dL
    HCT < 2.0 0-67% 0-67%
    PLT (whole blood) < 5.0 0-1900 x 10³/µL 1900-3800 x 10³/µL
    PLT (PLT concentrate) < 5.0 0-2800 x 10³/µL 2800-5600 x 10³/µL

    *CDR: Customised Dilution Ratio